Healthcare Professionals – Please note these additional Limitations
7. Patients undergoing oxygen therapy may have inaccurate results.
8. If the patient has the following conditions, the result may fail:
♦ Severe dehydration
♦ Severe hypotension (low blood pressure)
♦ Shock
♦ A state of hypoglycemic-hyperosmolar state (with or without ketosis)
9. Lipemic samples: Cholesterol level up to 500 mg/dL and triglycerides up to 3,000 mg/dL do not
affect the results. Grossly lipemic patient samples have not been tested and are not
recommended for testing with EasyPlus mini 2IN1 Blood Glucose Meter.
10. Critically ill patients should not be tested with home-use blood glucose meters.
11. Interfering Substances depend on the concentration. The below substances up to the test
concentration will not affect the test results.