The lead from the internal measuring circuitry of the meter to the LO terminal (on the
front panel) is the same for both voltage and current measurements. The resistance of this
lead is approximately .003 ohm. If current is being measured, therefore, a voltage drop
will occur in the resistance that is common to both circuits. This internal resistance, when
added to the external resistance of the lead from the COM input terminal will affect the
accuracy of the voltage reading. For instance, if the external lead resistance is .007 ohm,
the "total" common resistance is .010 ohm. If there is 1 A of current, the voltage reading
would be affected by:
(1 A x.01ohm ) = .01 V or 10 m V.
Depending on the circumstances, this may be significant.
If you want to measure dc voltage on an input signal in the primary display and dc
current in the secondary display, proceed as follows:
1. Turn the Meter on.
2. Press D to select the dc voltage function for the primary display.
3. Press Q then E to select the dc current function for the secondary display.
4. Connect the leads to the test circuit as shown in Figure A-2 and read the
measurements on the display. Although current will be displayed as negative, it is in
fact positive when interpreted according to current flow convention.
Response Times
Response time is the time between a change in an input and when that change is
displayed. The meter's response time depends on many factors: the measurement
A-4
Shop for Fluke products online at:
Table A-2. Sample Dual Display Applications (cont.)
Secondary
Display
• Select and sort resistors. (See also "Using the Compare Function"
Resistance
in Chapter 3.)
• Show actual measurement while holding a previous, stable
Actual Value
measurement on the primary display
Figure A-2. DC Voltage and DC Current Measurement on Input Signal
MyFlukeStore
www.
Applications
8808A
5-1/2 DIGIT MULTIMETER
INPUT
SENSE
V
4 W
2W/4W
HI
HI
1000V
1000 V
CAT I
MAX
300V
600 V
CAT II
750V
1V
LO
LO
500 V pk
200 mA
MAX
10 A
MAX
mA
10A
.ca
eue27.eps
1.877.766.5412

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents