Fluke 8808A User Manual page 82

Hide thumbs Also See for 8808A:
Table of Contents

Advertisement

8808A
Users Manual
Table 4-11. Range and Measurement Rate Commands and Queries (cont.)
Command
RANGE <value range>
RANGE1?
RANGE2?
RATE <speed>
RATE?
4-22
Shop for Fluke products online at:
Sets the primary display to <value range> where <value range> is the number
in the Range Value column of Table 4-11A that corresponds with the
applicable function ranges (voltage, ohms, current, etc.).
Table 4-11A. Ranges for Each Function
Range
Voltage
Ohms
Value
Range
Range
1
200 mV
200 e
2
2 V
2 ke
3
20 V
20 ke
4
200 V
200 ke
[1]
5
1000 V dc
2 Me
6
NA
20 Me
7
NA
100 Me
[1] 1000 V dc, 750 V ac
Returns the range presently selected on the primary display.
Returns the range presently selected on the secondary display. If the
secondary display is inactive, an Execution Error is generated.
Sets the measurement rate to <speed> where <speed> is either S for slow
(2.5 readings/second), M for medium (20 readings/second), or F for fast
(100 readings/second).
S, M, and F can be entered as uppercase or lowercase. Any other entry for
<speed> generates an Execution Error.
Returns <speed> as S for slow (2.5 readings/second), M for medium
(20 readings/second), or F for fast (100 readings/second).
MyFlukeStore
www.
Description
AC
Freq.
Current
Range
20 mA
2 kHz
200 mA
20 kHz
2 A
200 kHz
10 A
1000 kHz
NA
NA
NA
NA
NA
NA
1.877.766.5412
.ca
DC
Current
200 µA
2000 µA
20 mA
200 mA
2 A
10 A
NA