Weinmann prisma20C Instructions For Use Manual page 12

Sleep therapy devices, type wm 100 td and type wm 100 th health care product
Table of Contents

Advertisement

2 Safety
2.1.2 Power supply
Caution
2.1.3 Transport
Notice
12
EN
WM 100 TD
Risk of infection when reusing therapy device!
When the therapy device is used by multiple patients, infections
may be passed on to the next patient.
 Use a bacteria filter. When the device is used without a bacteria
filter: Have the device hygienically prepared by the
manufacturer, Weinmann, or an authorized dealer.
Treatment prevented due to increased resistance when
bacteria filters are used!
Misting or moistening can increase the resistance of the bacteria
filters, thereby modifying the output of the therapy pressure.
 Check bacteria filters for increased resistance and blockages
regularly and remove them.
Risk of injury due to inaccessible power plug!
An obstructed power plug cannot be pulled out in an emergency
and can thus result in injury.
 Keep the power plug and power supply accessible at all times.
Risk of injury and material damage as a result of insufficient
power supply!
Operation of the device outside the specified power supply range
can injure the user and damage the device.
 Only operate the device with the supplied power supply unit at
voltages from 100 V to 240 V.
 Use the DC adapter for operation at voltages from 12 V or 24 V.
Water in the device can cause material damages!
If the device is tilted severely, the residual water from the
respiratory air humidifier can enter the device and damage it.
 Do not transport or tilt the device when the respiratory air
humidifier is filled.
Dirt in the device can cause material damage!
Dirt entering the device during transport can damage the device.
 Only transport the device with the cover in position.
 Transport the device in the corresponding transport bag.

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Weinmann prisma20C

Table of Contents