Disinfecting
In the procedures below, only one disinfection process needs to be performed.
High level thermal
Part
Cleanable water tub
Air outlet
1. Soak the disassembled components in a hot water bath at 194ºF (90ºC) for 1 minute.
Take care that no air bubbles are trapped against the components.
2. Air dry out of direct sunlight and/or heat.
High level chemical
Part
Cleanable humidifier
Air outlet
1. Soak the disassembled components in a commercially available solution of a chemical sterilant.
Take care that no air bubbles are trapped against the components.
2. Thoroughly rinse the cleanable humidifier in drinking quality water (five liters per assembly) by immersing it completely for a
minimum of one minute in duration.
3. Repeat the rinse procedure two additional times using fresh water for a total of three rinses.
4. Air dry out of direct sunlight and/or heat.
Inspecting
Perform a visual inspection of all components. If any visible deterioration is apparent (cracking, crazing, tears, etc), the water
tub should be discarded and replaced. Slight discoloration of the silicone components may occur and is acceptable.
Validated number of cycles
Hot water: 194°F (90°C) for 1 minute.
130
130
Validated number of cycles
CIDEX OPA Ortho-phthalaldehyde
0.55% for 12 minutes
Gigasept FF (new)
5% for 15 minutes
130
130