ResMed AirCurve 10 VAuto User Manual page 15

Hide thumbs Also See for AirCurve 10 VAuto:
Table of Contents

Advertisement

 
Notes:
 The manufacturer reserves the right to change these specifications without notice.
 The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
 Do not use electrically conductive or antistatic air tubing.
 The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Pressure sensor at air outlet:
Mask pressure
Flow derived values:
Leak
Tidal volume
Respiratory rate
Minute ventilation
Ti
I:E ratio
Value
1
Pressure measurement
:
2
Mask pressure
Flow and flow derived values
Flow
2
Leak
2,3
Tidal volume
2,3
Respiratory rate
2,3
Minute ventilation
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
3
Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using
nominal ResMed mask vent flows.
Pressure accuracy
Maximum static pressure variation at 10 cm H
Without humidification
With humidification
Maximum dynamic pressure variation according to EN ISO 17510-1:2009
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure [cm H
O (hPa)]
2
4
8
12
16
20
25
14
Range
2–25 cm H
0 (2–25 hPa)
2
0–120 L/min
0–4000 mL
0–50 BPM
0–30 L/min
0.1–4.0 sec
1:100 – 2:1
1
Accuracy
±[0.5 cm H
0 (0.5 hPa) + 4% of measured value]
2
1
:
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
±20%
±1.0 BPM
±20%
O (10 hPa) according to EN ISO 17510-1:2009
2
Standard air tubing
± 0.5 cm H
O (± 0.5 hPa)
2
± 0.5 cm H
O (± 0.5 hPa)
2
10 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.3 / 0.3
Display resolution
0.1 cm H
1 L/min
1 mL
1 BPM
0.1 L/min
0.1 sec
0.1
SlimLine air tubing
± 0.5 cm H
± 0.5 cm H
15 BPM
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.5
0.5 / 0.4
O
2
O (± 0.5 hPa)
2
O (± 0.5 hPa)
2
20 BPM
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.8 / 0.8
0.7 / 0.7