For the customers in the U.S.A and Canada
When you use this product connected to 240 V single
phase, be sure to connect this product to a center tapped
circuit.
For the customers in Europe
This product has been manufactured by or on behalf of
Sony Corporation, 1-7-1 Konan Minato-ku Tokyo, 108-
0075 Japan. Inquiries related to product compliance
based on European Union legislation shall be addressed
to the authorized representative, Sony Deutschland
GmbH, Hedelfinger Strasse 61, 70327 Stuttgart,
Germany. For any service or guarantee matters, please
refer to the addresses provided in the separate service or
guarantee documents.
Important safeguards/notices for use in the
medical environments
1. All the equipments connected to this unit shall be
certified according to Standard IEC60601-1,
IEC60950-1, IEC60065 or other IEC/ISO Standards
applicable to the equipments.
2. Furthermore all configurations shall comply with the
system standard IEC60601-1-1.
Everybody who connects additional equipment to
the signal input part or signal output part configures
a medical system, and is therefore, responsible that
the system complies with the requirements of the
system standard IEC60601-1-1.
If in doubt, consult the qualified service personnel.
3. The leakage current could increase when connected
to other equipment.
4. For this particular equipment, all accessory
equipment connected as noted above, must be
connected to mains via an additional isolation
transformer conforming with the construction
requirements of IEC60601-1 and providing at least
Basic Insulation.
5. This equipment generates, uses, and can radiate radio
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference to other equipment. If this unit causes
interference (which can be determined by
unplugging the power cord from the unit), try these
measures: Relocate the unit with respect to the
susceptible equipment. Plug this unit and the
susceptible equipment into different branch circuit.
Consult your dealer. (According to standard EN60601-
1-2 and CISPR11, Class B, Group 1)
6. Model LMD-1951MD is a monitor intended for use
in a medical environment to display pictures from
cameras or other systems, other than diagnostic X-
ray equipment.
WARNING on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each
country if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power
Cord / Appliance Connector / Plug, please consult a
qualified service personnel.
WARNING on power connection for medical
use
Please use the following power supply cord. With
connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord
that is approved for use in your area.
United States and Canada
Plug Type
HOSPITAL GRADE*
Cord Type
Min. Type SJT
Min. 18 AWG
Minimum Rating
10A/125V
for Plug and
Appliance
Couplers
Safety Approval
UL Listed and CSA
*Note: Grounding reliability can only be achieved when the
equipment is connected to an equivalent receptacle marked 'Hospital
Only' or 'Hospital Grade'.
Symbols on the unit
Symbol
Location
Bottom
Bottom
Rear
Rear
Front
-
This symbol indicates
Power switch.
Press to turn the monitor
on.
Power switch.
Press to turn the monitor
off.
The equipotential
terminal which brings the
various parts of a system
to the same potential.
Functional earth terminal
Key inhibit
The setting are locked so
that they cannot be
changed.
3