Assembly Procedure - Nikon Eclipse 50i POL Instruction Manual

Polarizing microscope
Table of Contents

Advertisement

Assembly Procedure

4.2
1
Checking the input voltage
Make sure that the input voltage indicated on the back
panel of the microscope is the same as the voltage
provided in your area.
CAUTION
If the indicated voltage is different, do not
use the microscope. Contact your nearest
Nikon representative.
2
Attaching the circular graduated stage
(1) Remove the cushioning material from the substage
section and turn the coarse focus handle until the
elevating section is brought to the lowermost
position.
(2) Loosen the clamp screw for the circular graduated
stage.
(3) Place the circular graduated stage on the substage.
The stage must be level when fixed and its clamp
screw must be located facing the front side.
(4) If necessary, attach the specimen holders into the
hole on the stage.
For information about the optional attachable
mechanical stage, see "3.5.2 Attachable Mechanical
Stage (Option)."
3
Attaching the condenser
To perform the polarization microscopy, the P swing-out
condenser must be attached.
(1) Turn the coarse focus knob until the elevating
section is raised to the uppermost position.
(2) Turn the condenser focus knob until the substage is
brought to the lowermost position.
(3) Insert the condenser and adjust so that it faces
toward the front. Secure in place with the tool
stored in the back of the microscope.
(4) Turn the condenser focus knob until the substage is
raised to the uppermost position.
Chapter 4 Assembly
57
Specimen holder
Specimen holder
Condenser focus knob
Clamp screw for the condenser
Input
voltage
indication
Clamp screw
for the stage