Advertisement

• testing control solutions at high altitude
without applying correction factors (read
the SureStepPro Glucose Control Solution
package insert for more information)
CAUTION
If the control result falls outside the expected
range, repeat the test. The system may not
be functioning properly. Do not use the
system to test patient blood glucose levels
unless the control solution result falls within
the expected range.
If you have problems using the meter to test
the control solutions, or if the meter
malfunctions, call the LifeScan Healthcare
Professional Line at 1 800 524-7226.

Linearity Test

A Linearity Solution kit is available from
LifeScan. The kit contains five levels of
glucose solution used to check the linearity
of the system.
Refer to your institution's policy for
establishing the linearity of a test method.