FilterSpecifications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
MaximumFlowrate
Measured
pressure at the
patient connection
port (cmH20)
Average flow at
the patient
connection port
(l/min)
Long Term Static Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.5 cmH2O
DynamicShort-TermPressureAccuracywithandwithout
DV5HHHumidifier
Waveform
Sinusoidal cycle with
I:E = 1:1
Equipment classification with respect to protection from electric shock . . . . . . . . . . .Class II
Degree of protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Part
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX1
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air
or with oxygen or nitrous oxide.
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
SE-DV53-1
30
(perISO17510-1:2007)
3.0cmH20 7.0cmH20 12.0cmH20 16.0cmH20 20.0cmH20
2.9
6.4
83.8
126.8
(perISO17510-1:2007)
Volume
BreathRate
[mL]
[min-1]
500
10, 15, 20
SpEcificationS
TestPressures
11.0
15.0
137.4
130.0
SetPressure
[cmH2O]
3.0
7.0
12.0
16.0
20.0
19.0
113.8
PressureAccuracy
[Pk-PkcmH2O]
0.5
0.5
1.0
1.0
1.0