Invacare Wheelchair Pronto M91 Base Owner's Manual page 29

Base with surestep
Hide thumbs Also See for Wheelchair Pronto M91 Base:
Table of Contents

Advertisement

Information Gauge Display Diagnostics
DISPLAY
Information
Gauge
Display
Service Indicator Light Diagnostics
NUMBER
ERROR CODE
OF
DESCRIPTION
FLASHES
1
User Fault
2
Battery Fault
3
Left Motor Fault
4
Right Motor Fault
5
Left Park Brake Fault
6
Right Park Brake Fault Contact Invacare/Dealer for service.
7
Remote Fault
8
Controller Fault
9
Communications Fault Contact Invacare/Dealer for service.
10
General Fault
11
Incompatible or
incorrect Remote
Part No 1143153
SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
DESCRIPTION
All LEDs are off.
All LEDs are on.
Left RED LED is flashing.
Left to Right "chase"
alternating with steady
display.
All LEDs are flashing slowly. Joystick has detected Out-
Release joystick to neutral and try again.
Charge the batteries. Refer to Charging Batteries on page 63.
Check that battery cables are connected properly. Refer to Connecting/
Disconnecting the Battery Wiring Harness on page 60.
If necessary, replace batteries. Refer to Installing/Removing the Batteries
on page 57.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Check to make sure joystick is connected properly.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Wrong type of remote connected. Contact Invacare/Dealer for service.
29
DEFINITION
Power is off.
Power is on.
Battery charge is low.
Joystick is in
programming, inhibit and/
or charging mode.
of-Neutral-at-Power-Up
mode.
POSSIBLE SOLUTION
Pronto® M91™Base with SureStep®
COMMENTS
Fewer than three LEDs on
implies reduced battery
charge.
The batteries should be
charged as soon as
possible.
The steady LEDs indicate
the current state of the
battery charge.
Release the joystick back
to Neutral.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents