Chair adjustments
Angle adjustment
Place the chair on a flat surface
(such as a table). Loosen the two
eccentric bolts (1) and 10mm A/F
nuts (2) that secure the castor
assembly (3) to the frame (4). See
Fig. 50. Follow the procedure
detailed on page 34 for angle
adjustment. (Fixed front frame).
Inside mounted
To inside mount castors, reverse
positions of castor assemblies.
Remove castor assemblies from
frame (noting right positioned
castor and left positioned castor).
Inside mount right castor on the left
side and inside mount the left
castor on the right side.
Forward or backward placement
Loosen the two eccentric bolts (1).
Remove the clevis pin (2) by taking
off the locking spring (3). Slide the
castor housing (4) forward or
rearward and insert clevis pin into
one of three pre-drilled frame holes
(5). Follow the procedure for angle
adjustment given above. Repeat
the process for the other castor
(Fig. 51). (Fixed front frame).
2 Issue 4
3
1
Fig. 50
4
1
2
Fig. 51
36
2
4
90°
3
5
90°