Test Lead Alert; General Specifications - Fluke 175 Calibration Information Manual

True rms multimeters
Table of Contents

Advertisement

Test Lead Alert

Personal injury or damage to the Meter can occur if you attempt to make a
measurement with a lead in an incorrect terminal.
To remind you to check that the test leads are in the correct terminals, LEAd is displayed when the rotary
switch is moved to or from any
Specifications

General Specifications

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity
at 0 % to 75 %. Accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] ). AC conversions are
ac-coupled, average responding, and calibrated to the rms value of a sine wave input.
Maximum voltage between
any terminal and earth ground: 1000 V DC or AC RMS
Surge Protection:
W Fuse for mA inputs:
W Fuse for A input:
Display:
Altitude:
Temperature:
Temperature coefficient:
Electromagnetic Compatibility
(EN 61326-1:1997):
Relative Humidity:
Relative Humidity in
50 MΩ Range:
Battery Life:
Size (H x W x L):
Weight:
Safety Compliances:
Certifications:
WWarning
position.
A
8 kV peak per IEC 61010
440 mA, 1000 V FAST Fuse, minimum interrupt rating 10 kA
11 A, 1000 V FAST Fuse, minimum interrupt rating 17 kA
Digital: 6000 counts, updates 4/sec
Bar Graph: 33 segments;
Updates 40/sec
Frequency: 10,000 counts
Capacitance: 1,000 counts
Operating: 2000 m; Storage: 12,000 m
Operating: −10 °C to +50 °C;
Storage: −30 °C to +60 °C
0.1 X (specified accuracy / °C
(< 18 °C or > 28 °C)
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature:
specified accuracy ± 5 °C (9 °F)
0 % to 90 % @ 0 °C to 35 °C;
0 % to 70 % @ 36 °C to 50 °C
0 % to 80 % @ 0°C to 35 °C;
0 % to 70 % @ 36 °C to 50 °C
Alkaline: ~200 hrs typical
4.3 cm x 9 cm x 18.5 cm
420 g
ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage
Category III, 600 V Overvoltage Category IV
CAT III- Equipment designed to protect against transients in equipment in fixed-
equipment installations, such as distribution panels, feeders and short branch
circuits, and lighting systems in large buildings.
CAT IV- Equipment designed to protect user against transients from the primary
supply level, such as an electricity meter or an overhead or underground utility
service.
CSA, TÜV (EN61010), P, ; (N10140), UL (3111), VDE
True RMS Multimeters
Test Lead Alert
3