Environmental Requirements; Temperature; Relative Humidity; Altitude - Kodak DryView 6800 User Manual

Kodak dryview 6800 laser imager user's guide
Hide thumbs Also See for DryView 6800:
Table of Contents

Advertisement

Specifications

Environmental Requirements

Temperature

Relative Humidity

Altitude

Floor Vibration

6-2
IMPORTANT:
• This equipment is intended to connect to other medical devices. Only an
authorized Service Provider of Kodak Products or Customer Qualified
Service Personnel may install this equipment.
• Do not locate the Imager within 2 m (6.56 ft) of where a cell phone will
be used, even if the cell phone is separated from the equipment by a
wall.
• Do not locate the Imager within 4 m (13.12 ft) of a microwave oven.
Electromagnetic radiation from a microwave oven is only an issue if,
after the oven door is closed and latched, the seal does not maintain an
electromagnetic tight fit between the oven door and the oven main
housing. Determining if the seal has an electromagnetic tight fit requires
special detection equipment.
• Do not locate the Imager in the presence of flammable anesthetics,
oxygen, or nitrous oxide. The Imager does not have a gas-sealed
electronics enclosure and could ignite any flammable or explosive gases
present in the Imager's environment.
• Do not locate the Imager in close contact with MRI devices due to
possible high magnetic fields near an MRI unit. The magnetic field in the
area where the Imager is installed must be less than 50 Gauss.
• Do not locate the Imager closer than 1.83 m (6 ft) from a patient bed or
chair.
• Do not substitute or modify any part of the Imager without prior written
approval of Eastman Kodak Company.
• Operating: 15° to 35° C (59° to 95° F)
• Storage: -35° to 60° C (-31° to 140° F)
• Operating: 20% to 85% RH, noncondensing
• Storage: 10% to 90% RH, noncondensing
• 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
For the Imager to operate with no image degradation, floor vibration levels
cannot exceed the following.
2
G
/Hz
Freq. Range
-7
1 x 10
1-50 Hz
8F927
Direction
Vertical or horizontal
January 17, 2007