Replacing The Leg Actuators - Invacare ROZE Owner's Manual

Stand up patient lift
Hide thumbs Also See for ROZE:
Table of Contents

Advertisement

Available languages

Available languages

SECTION 7—MAINTENANCE
Washer
TOP
Cap
Screw

Replacing the Leg Actuators

NOTE: For this procedure, refer to FIGURE 7.2 on page 25.
NOTE: If the legs of the patient lift do not open and close properly, it may be necessary to replace one or both of the leg actuators. 
1.
Test the operation of the legs:
Press the CLOSE button on the hand control to close the legs.
Press the OPEN button on the hand control to open the legs.
2.
While holding on to the mast handles, carefully tip the patient lift back until it rests on the floor.
NOTE: It is necessary to remove both small base covers before removing the large base cover.
3.
Remove the sixteen screws that attach the small base covers to the base.
4.
Remove the eight pan head screws that attach the large base cover to the base.
NOTE: If the left or the right leg of the patient lift does not operate properly, perform the following steps on the side of the base necessary 
to replace the actuator. If both legs do not operate properly, perform this replacement procedure on both sides of the base.
5.
Locate and disconnect the leg actuator lead wires. 
Roze™ Stand Up Patient Lift
Actuator Cap
Nylon
Washers
Bearing
Bushing
Actuator Lead
Routing Hole
Bushing
Washer
Cap
Bearing
BOTTOM
Screw
Mast Assembly
Actuator Mounting
FIGURE 7.1 Replacing the Electric Actuator
Flat Washer
Washer
Bearing
Lift Arm Actuator
Mounting Bracket
Flat
Washer
Nylon
Washers
Cap
Bearing
Nut
Washer
Bracket
24
Cap
Nut
Electric Actuator
Part No. 1150703

Advertisement