• If a vertical calibration fails and the module is removed before dismissing the
• Timebase resolution changed from 3 digits to 4, giving the same behavior as the
• Remotely invoking an extinction ratio cal, where a vertical cal is required, brings
• Remotely sending the TEDGE command caused various problems. The fix is to
• Remotely selecting vertical markers from a time waveform to a versus waveform
• Invalid dates remotely sent to the instrument cause several problems, which are
N O T E
• None
• User Vertical Calibration routine that causes vertical calibration failures
dismiss the Message Box. The fix was to re- enable the touchscreen before
displaying the Message box and the disabling it after it has been dismissed.
Message Box an illegal operation error occurs. The Fix is to check for the presence
of the module before reading from it.
83480A.
up a dialog box that cannot be cleared. The dialog box is replaced with a status
message in the status bar.
only allow TEDGE? query and to reject TEDGE command.
caused a "data out of range" error message; this has been fixed.
now resolved.
A.01.13 Release Notes, May 2, 2000 (Differences from Release
A.01.12)
This firmware release is not supported.
New Feature List
Defects Fixed
N1010A FlexDCA and 86100A/B/C/D
Firmware Release Notes
47