Release Notes, July 18, 2002 (Differences From Release A.03.03) - Agilent Technologies 86100A Release Notes

Agilent 86100a/b/c/d infiniium dca series & n1010a flexdca, firmware release notes (a.12.02 dca and below & a.02.50 flexdca and below)
Hide thumbs Also See for 86100A:
Table of Contents

Advertisement

N1010A FlexDCA and 86100A/B/C/D
Firmware Release Notes
• Pressing Default Setup with no modules was displaying "Entry was not a valid
selection" in the Dynamic Text Area
• Corrected problem with software upgrade process, cycling power too quickly could
prevent upgrade from completing
A.03.03 Release Notes, July 18, 2002 (Differences from Release
A.03.03)
New Feature List
• Add warning when attempting to save a file to a full floppy disk
• Add RZ/NRZ Eye mode selection GUI button
• Renamed horizontal and vertical calibration dialog box to avoid confusion with
the precision timebase module
• TDR Differential Normalization feature added
• New masks: Infiniband, 10 GbE, Xaui
• Changes to electrical channel offset DAC for 86101A and 86103A modules
• The following new modules are supported: 86105B, 86107A Opt 010, 86117A/B,
86118A
Defects Fixed
• Modified the RZ algorithm to take care of unsymmetrical edges
• Corrected a problem with mask file names with a zero instead of an ?gO?h i.e.
STM64_OC192 written as STM64_0C192?h
• Acquisition Limits misbehave with long patterns is now fixed