Automatic Trigger Options

In the trigger menu, settings for automatic triggering can
be changed as follows. (See also Chapter 1: "Displaying
an Unknown Signal with Connect-and-View")
1
Display the
The
key labels can differ depending on
TRIGGER
the latest trigger function used.
2
Open the Trigger Options menu.
3
Open the Automatic Trigger menu.
54
key labels.
TRIGGER
Note
If the frequency range of the automatic triggering is set to
> 15 Hz, the Connect-and-View™ function responds more
quickly. The response is quicker because the test tool is
instructed not to analyze low frequency signal
components. However, when you measure frequencies
lower than 15 Hz, the test tool must be instructed to
analyze low frequency components for automatic
triggering:
4
Select > 1 H
measurement screen.
and return to the
Z