Fluke 192B - 196B/C - 199B/C
Users Manual
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges .............................same as VDC, VAC, VAC+DC
Probe Sensitivity ................. 100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy...........................same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)

Recorder

TrendPlot (Meter or Scope)
Chart recorder that plots a graph of min and max values of
Meter or Scope measurements over time.
Measurement Speed .................. 5 measurements/s max
Time/Div ........................................... 5 s/div to 30 min/div
Record Size ............................................... ≥18000 points
Recorded Time Span... 60 min to 22 days(single reading)
30 min to 11 days (dual reading)
Time Reference ..................... time from start, time of day
104
Scope Record
Records scope waveforms in deep memory while
displaying the waveform in Roll mode.
Source...................................................... Input A, Input B
Max. Sample Speed (5 ms/div to 1 min/div) .........20 MS/s
Glitch capture (5 ms/div to 1 min/div) ....................... 50 ns
Time/Div in normal mode ................. 5 ms/div to 2 min/div
Record Size .................................. 27000 points per input
Recorded Time Span ................................ 6 s to 48 hours
Acquisition Modes ....................................... Single Sweep
Time Reference ..................... time from start, time of day
Continuous Roll
External Triggering