Control Loop
After turning the function rotary switch, the voltage display mode is reached.
Pushing "DISPLAY MENU" now calls up the control loop. According to the
selected measuring function, different setting values can be displayed and
changed in the control loop. The "DISPLAY MENU" button switches between
the different set values inside a contiuous loop. The "SELECT" button selects
the decimal point to be changed. Pushing the "CHANGE ITEM" button the
instrument either switches between certain set values or increases the decimal
point selected with "SELECT" by 1.
After parmeter setting has been finished the next display can be called with
"DISPLAY MENU" or the measurement can be started with "START TEST".
Depending on the selected function, the following parameters can be displayed
or changed:
Function
Parameter
U ST
F ST
R E
F M
3pole
U M
and
R K
R E LIMIT
}
R E
(warning
4pole
Sound)
R*
R E
U ST
F ST
28
Setting Range
(AFC/94/105/111/128) Hz
48 V/20 V
0.000
... 29.99
0.000
... 999 k
On/Off
On/Off
Remarks
display only
display only
selectable to 20 V
with CODE
in position R E
3pole only *
only if activated
with CODE
only if R E LIMIT
is activated with
CODE
only if activated
with CODE
display only
display only