Fluke 1625 User Manual page 19

Earth/ground tester
Hide thumbs Also See for 1625:
Table of Contents

Advertisement

Operation:
Working temperature range:
Operating temperature range:
Nominal temperature range:
Storage temperature range:
The chart of four temperature ranges for the instrument exist to sat-
isfy European Standards requirements; the instrument can be used
over the full Working temperature range by using the temperature co-
efficient to calculate accuracy at the ambient temperature of use.
Temperature coefficient:
Operating errors:
The maximum percentage operating error within the measurement range does
not exceed ± 30 % with the measured value as fiducial value, as determined in
accordance with TABLE 1.
The operating error applies under the rated operating conditions given in
IEC1557-1 and the following:
injection of series interference voltages with system frequencies of
400 Hz, 60 Hz, 50 Hz, 16
across the terminals E (ES) and S. The r.m.s. value of the series inter-
ference voltage shall be 3 V;
resistance of the auxiliary earth electrode and of the probes: 0 to 100 x
R
but 50 k ;
A
system voltages between 85 % and 110 % of the nominal voltage and
between 99 % and 101 % of the nominal system frequency for meas-
uring equipment with a mains supply and/or measuring equipment de-
riving its output voltage directly from the distribution system.
Limits of error:
Climate class:
Type of protection:
Central rotary switch and function keys
-10 °C
0 °C
18 °C
-30 °C
Note
± 0.1 % of range / Kelvin
refer to operating temperature range and RH
< 20 RE, RS < 100 RE
2
/
Hz or with d.c. voltage respectively
3
refer to nominal temperature range
C1 (IEC 654-1), -5 °C...+45 °C, 5 %...95 % RH
IP 56 for case, IP 40 for battery door according to
EN 60529
Earth/Ground Tester
... +50 °C
... +35 °C
... +28 °C
... +60 °C
Specifications
11