Agilent Technologies 34980A User Manual page 280

Multifunction switch/measure unit
Hide thumbs Also See for 34980A:
Table of Contents

Advertisement

10
64-Bit Digital I/O Module with Memory and Counter
To read digital data with more control over the channel parameters,
use the SCPI
set up the digital I/O channel parameters. For example, sending the
following SCPI command to a Digital I/O module installed in slot 1 of the
mainframe, sets a 16- bit input channel (103) to use a 2.5 V input
threshold, and normal polarity.
CONF:DIG WORD, 2.5, NORM, (@1103)
Once configured, the data is read using the following command.
SENS:DIG:DATA:WORD? (@1103)
You may also read an individual bit using the
This allows you to check the state of an individual bit in a channel
without having to create an input mask. For example, the following
command returns the state of bit 3 in the channel 101 byte.
SENS:DIG:DATA:BIT? 3, (@1101)
The acceptable range for the bit parameter is based on the channel width
as shown below:
• BYTE (8- bit): <bit> can range from '0' to '7'
• WORD (16- bit): <bit> can range from '0' to '15'
• LWORd (32- bit): <bit> can range from '0' to '31'
The
change the direction (input or output) of the channel. If the channel is
configured as an output, the
driven.
Writing Digital Data
To write digital data, set the channel output parameters using the
commands. For example, sending the following SCPI commands to a Digital
I/O module in slot 1 sets a 32- bit channel to use normal polarity,
with active drive and a 'set' output voltage of 4 volts.
CONF:DIG:WIDT LWOR,(@1201)
CONF:DIG:POL NORM,(@1201)
SOUR:DIG:DRIV ACT,(@1201)
SOUR:DIG:LEV 4,(@1201)
The width and polarity parameters apply to both input and output
operations.
268
CONFigure
command differs from the
SENSe
and
commands. The
SENSe
MEASure
command will return the value being
SENSe
commands
CONFigure
commands.
SENSe
command in that it will not
34980A User's Guide