Using Single Drive Switches and Attenuators

Some microwave switches require a single drive. With single drive devices
the channel numbering is not consecutive across all channels in a bank
(refer to the channel numbering description on
The 34945A can provide single drive devices with either pulsed or
continuous drive current. Settings and parameters for continuous drive
mode are given in the next section.
Continuous Drive Current
Driving non- latching devices requires a power supply capable of handling
sustained high current requirements. You may only use continuous drive
current with channel configured for single drive. Additionally, to prevent
power supply loading, care must be taken when operating more than one
continuous drive at a time. The actual drive may be configured as either
TTL or open- collector operation.
Using Continuous Drive
The diagram below illustrates the continuous drive signals for two
channels (switches) and the relationship of the drive parameters to the
power supply requirements.
As shown in the diagram, the drive signal is initially applied to channel 1.
Drive is applied to channel 2 only after a power supply recovery period
has elapsed T(Recovery). The power supply recovery time is set using the
the
be set individually for each channel or will default to 0.0 ms following
either a
command.
If you are verifying the channel closure (see
specify a T(Settle) parameter. This parameter ensures the switch has had
time to change state before the position indicator is evaluated. This
parameter may be set individually for each channel or will default to
0.0 ms following either a
ROUTe:RMODule:BANK:PRESet
34980A User's Guide
Drive Ch 1
Drive Ch2
Start Drive
Channel 1
ROUTe:CHANnel:DRIVe:TIMe:RECovery
SYSTem:RMODule:RESet
Microwave Switch/Attenuator Driver
T(Setttle)
T(Recovery)
Start Drive
Channel 2
command. This parameter may
or
ROUTe:RMODule:BANK:PRESet
page
SYSTem:RMODule:PRESet
command.
page
212).
Channel 1 Position
Indicators Evaluated
218), you may also
or
8
215