The instrument can be optionally equipped with an output/power supply board, to allow
for the following capabilities:
1. RS-232C.
2. Current Loop output (4-20mA or 0-24mA).
3. Operation from an external power supply e.g. a wall transformer.
6.1 RS-232C
The RS-232C output is provided on a 9 pin sub "D" female connector, located at the rear
of the instrument. The configuration is 9600 baud, Even Parity, 7 Bits, 1 Stop. The inter-
face is isolated from the sensor and case (however as a sefety precoution there is a 1M
shunt to prevent electrostatic buildup and discharge), it is configured as a modem, thus it
can be directly plugged into a PC. To also use the current loop output, a special cable is
needed since the connector is shared, see the pin out table below.
In order to receive the dewpoint in the currently selected engineering units, the host com-
puter must query the 8800P with:
The 8800P will respond with the dewpoint, followed by characters indicating the currently
selected engineering units, followed by carriage return e.g.:
If an error condition exists the 8800P will respond to the question mark with a phrase
describing the error:
SensOpen
NOTE: Characters marked in bold are indications of sent or received strings.
pin
Signal
1
Current Loop out
2
3
4
strapped to 6
5
Gnd-RS232 & Current Loop
-79.1degC
-110.4degF
0.0339LbsH2O/mmscf
0.625ppmV
0.00544g/m3
C
or SensSaturated
R
9 pin "D" female connector pin assignments
pin
6
Tx
7
Rx
8
9
page 11
?
C
R
C
R
or
C
R
or
C
R
or
C
R
or
C
R
C
or SensShorted
R
strapped to 4
looped from pin 8 through RS-232 level driver
looped to pin 7 through RS-232 level receiver.
current loop select:Open=0-24mA,Grounded=4-20mA
C
R
Signal