The Analyze Function - Teledyne 7320 Operating Instructions Manual

Infrared gas analyzer
Table of Contents

Advertisement

4 Operation
Ranges that overlap
Ranges whose limits are entirely within the span of an adjoining
range.
Figure 4-2 illustrates these schemes graphically.
Figure 4-2: Examples of Autoranging Schemes
4.7

The Analyze Function

Normally, all of the functions automatically switch back to the Analyze
function when they have completed their assigned operations. Pressing the
Escape button in many cases also switches the analyzer back to the Analyze
function. Alternatively, if you leave your analyzer on MAIN MENU screen
within 5 seconds without touching any key, it will automaticaly return to analyze
function.
If the analyzer is in subfunction mode, in most cases, it will
automaticaly return to analyze mode within 10 minutes.
The Analyze function screen shows the impurity concentration and the
application gases in the first line, and the range in the second line. In the lower
right corner, the abbreviation
mode. If there is an * before the
4-26
indicates that the analyzer is in the Analyze
, it indicates that the range is linearized.
indicates non inverting range
indicates inverting range
Teledyne Analytical Instruments
Model 7320

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents