Using The "1Man" Calibration Option; Zero & Span Calibrating A Single Sensor Ch. M - Teledyne 1220 Instruction Manual

Multi - channel combustible monitor
Table of Contents

Advertisement

4 Operation
The two following calibration procedure can be performed by one
technician.
1. At the Front Panel of the Channel Module to be calibrated:
2. At the remote probe site, introduce certified span gas to the sensor.
3. Wait 60 seconds so that the reading can stabilize.
4. Remove the span gas and introduce zero gas to the sensor.
5. Wait 60 seconds so that the reading can stabilize.
6. At the Front Panel of the Channel Module being calibrated:
NOTE: Information about calibration values for various combustible
NOTE: If you need to adjust the Sensitivity Gauge, see Paragraph 4.8
7. If necessary, proceed to the previous or next procedure. Otherwise,
1. At the Front Panel of the Channel Module to be calibrated:
4-10

4.4.5 Using the "1MAN" Calibration Option

4.4.5.1
Zero and Span Calibrating a Single Sensor
Channel Module
a. Press SELECT.
b. Use the right or left arrow key to highlight the 1MAN option.
c. Press ENTER.
a. Press the up or down arrow key until the display reads "0".
b. Press ENTER.
c. Press the up or down arrow key until the display reads the value
that corresponds to the value of the certified span gas used.
gases appears in Table 1 on page 2-5 and in Section 4.11 on page
4-14
for recommendations.
d. Press ENTER.
press RUN to place the instrument in the analysis mode.
4.4.5.2
Zero and Span Calibrating a Dual Sensor Channel
Module
a. Press SELECT.
b. Use the up or down arrow key to highlight the appropriate
sensor to be calibrated (S1 = Sensor 1 or S2 = Sensor 2).
c. Press ENTER.
d. Use the right or left arrow key to highlight the 1MAN option.
e. Press ENTER.
f. Perform the procedure listed in Paragraph 4.4.5.1 beginning
with Step "2" and then return here.
Teledyne Analytical Instruments
Model 1220

Advertisement

Table of Contents
loading

Table of Contents