Teledyne 3000ZA Operating Instructions Manual page 58

Trace oxygen analyzer
Table of Contents

Advertisement

Operation
back to non-alarm conditions. This mode requires an alarm to
be recognized before it can be reset. In the non-latching
mode, the alarm status will terminate when process
conditions revert to non-alarm conditions.
8. Are either of the alarms to be defeated?
9. The defeat alarm mode is incorporated into the alarm circuit
so that maintenance can be performed under conditions
which would normally activate the alarms.
10. The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in
Section 4.3.3 to enter your password. Once you have clearance to
proceed, enter the Alarm function.
Press the Alarm button on the front panel to enter the Alarm
function. Make sure that AL–1 is blinking.
Set up alarm 1 by moving the blinking over to AL–1 using the __
arrow keys. Then press Enter to move to the next screen.
Five parameters can be changed on this screen:
Value of the alarm setpoint,
value can be set from 0 to 10,000.00 ppm although the
sensor output is good only to 250 ppm
Out-of-range direction,
Defeated?
Failsafe?
Latching?
To define the setpoint, use the __ arrow keys to move the blinking
over to
. Then use the __ arrow keys to change the number.
AL–1 ####
Holding down the key speeds up the incrementing or decrementing.
(Remember, the setpoint units are ppm O
Teledyne Analytical Instruments
AL—1
AL—2
Choose Alarm
AL—1 10 ppm HI
Dft—N Fs—N Ltch—N
AL–1 ####
or
HI
LO
(Yes/No)
Dft–Y/N
(Yes/No)
Fs–Y/N
(Yes/No).
Ltch–Y/N
.)
2
3000ZA
ppm (oxygen);
46