Whenever the Instrument Event Register is read, the condition bits are cleared.
This register is used in conjunction with the Instrument Event Enable Register to determine the conditions
under which the Instrument Event Bit of the Status Byte is set.
Bit
0
1
2
3
4
5,6
7

Computer Interface Command Set

Generally, RS-232 and IEEE-488 commands are identical. A few exceptions apply only
when the RS-232 interface is active.
See Table 4-7 for a summary of computer interface commands and queries. A detailed
description of each command or query can be found in Table 4-8. (alphabetically
arranged.)
Computer interface command descriptions use angle brackets (< >) to
denote a parameter that must be supplied by the user or a string that is
returned by the instrument.
4-22
Table 4-6. Instrument Event Register (IER)
Name
ALT
Alarm Limit Transition. Set high (1) when any measurement value has
transitioned into or out of alarm. Alarms are defined through the front panel or
the computer interface (ALARM_LIMIT). This bit is cleared when alarms or
review values are cleared.
TOB
Totalize Overflow. Set high (1) when the Totalizer overflows (>65,535). This
bit is cleared when the Totalizer is zeroed from the front panel or set to some
other non-overflow value (<65,535) with the computer interface TOTAL
command.
OTC
Open Thermocouple. Set high (1) when open thermocouple checking is
enabled and any thermocouple channel is measured with a source
impedance greater than 1 to 4 kilohms.
CCB
Calibration Corrupted. Set high (1) when the instrument calibration data is
corrupted. When a calibration data check shows a corruption of calibration
data, the calibration alarm bit remains set in the Instrument Status Register
until the instrument is recalibrated.
CNC
Configuration Corrupted. The instrument configuration stored in NVRAM has
been found to be corrupted. The RAM CRC is no longer valid.
not used
SCB
Scan Complete. Set high (1) when a measurement scan has been completed.
Note
Description
Note