Mx+B Scaling

The menu for setting Mx+B scaling values takes you through the following steps:
Set the M value (sign and number).
Set the decimal point location for the M value.
Set the multiplier for the M value (m, x1, k, M).
Set the B value (sign and number).
Set the decimal point location for the B value.
Set the multiplier for the B value (m, x1, k, M).
Use
,
,
to accept your choice and advance to the next step.
For example, to set an M value of 1.8 and a B value of 32 for channel 7, do the
following:
1. Select channel 7 by pressing
2. Press
to begin Mx+B scaling setting for channel 7.
3. Use the arrow buttons to select "+18000". Press
Press
or
4. Use the arrow buttons to select the decimal point position ("+1.8000"). Then press
to accept this position.
5. Next, use the arrow buttons again to select the multiplier ("x1"). Press
accept this multiplier.
6. Use the arrow buttons to select +00320. Then press
7. Use the arrow buttons to select the decimal point position ("+0032.0"). Then press
to accept this position.
8. Next, use the arrow buttons again to select the multiplier ("x1").Press
accept this multiplier as the multiplier for the B value.
Mx+B scaling is now set at 1.8x+32 for channel 7.
Pressing
while partway through setting the M value undoes all changes entered thus
far, returning the instrument to Inactive Mode. Pressing
setting the B value cancels entries made thus far for the B value, but does not affect any
changes already made to the M value.

Setting the Scan Interval

Press
to set the scan interval. The latest value appears on the right display. The
format of the display is H:MM:SS. Use
to select the new value for a digit. Values ranging from 0:00:00 to 9:99:99 are
allowed. Press
Set the interval to 0:00:00 for continuous scanning. If you set the interval to a value
shorter than the time required to measure all channels, scanning effectively becomes
continuous.
and
to cycle through the selections at each step. Then press
or
to select the value for each digit. (Ignore the decimal point). Press
to accept these digits.
to accept the displayed value.
.
or
to move between digits.
while partway through
and
to move between digits. Use
Overview

Setting the Scan Interval

to
.
to
and
2
2-15