Table 4-8. Command and Query Reference (cont)
Remote with Lockout (RS-232 only)
Enter the IEEE-488.1 remote with front panel lockout (RWLS) state. All
front panel buttons are disabled, and the REM annunciator is lit.
If this command is used with the IEEE-488 interface, an Execution Error
is generated.
Set Mx+B Scaling Values
Set the M and B scaling values for the indicated channel, and display the
results of the Mx+B calculation in the indicated display range. Successful
execution of this command clears all
values in the Review array (all channels.)
SCALE_MB <channel>,<M_value>,<B_value>,<disp_range>
<channel> = (0 .. 20)
<M_value> = signed numeric quantity
<B_value> = signed numeric quan<disp_range> = (16)
<disp_range> = (1 .. 16)
CODE DISPLAY
MAX OFFSET CODE
OFFSET
1
0.0000 m
2
00.000 m
3
000.00 m
4
0000.0 m
5
0.0000 x1
6
00.000 x1
7
000.00 x1
8
0000.0 x1
When M=1 and B=0, Mx+B scaling is effectively nonexistent. The values
for M and B must be in the span +/-9999.9 Mega (E+6).
An Execution Error is generated for any of the following:
Invalid entries for channel number, M or B values, or display range code
are used.
The range required by the B value is larger than allowed by the display
range code.
The channel is defined as OFF.
Measurements are active.
Mx+B scaling values for a channel are automatically reset to 1 (M) and 0
(B) when the function for that channel is changed.
DISPLAY
VALUE
0.01
9
0.0000 k
0.1
10
00.000 k
1.0
11
000.00 k
1.0
12
0000.0 k
10.0
13
0.0000 M
100.0
14
00.000 M
1000.0
15
000.00 M
1000.0
16
0000.0 M
MAX
VALUE
1.0E4
1.0E5
1.0E6
1.0E6
1.0E7
1.0E8
1.0E9
1.0E10