Fluke 2620A User Manual page 100

Hydra series ii data acquisition unit, hydra series ii data logger
Hide thumbs Also See for 2620A:
Table of Contents

Advertisement

2620A, 2625A
Users Manual
ALARM_LIMIT
ALARM_LIMIT?
DATE
4-30
Table 4-8. Command and Query Reference (cont)
Set Alarm Limit
Store alarm limit information for the indicated channel and limit. The
fields to be given (in order) are:
ALARM_LIMIT <channel>,<limit_num>,<sense>,<value>
<channel>
= (0 .. 20)
<limit_num>
= 1 2
<sense>
= HI LO OFF
<value>
= Signed numeric quantity
A Command Error is generated if a value is supplied when the sense is
OFF. An Execution Error is generated under any of the following
circumstances:
<value> is outside the range +/-9999.9 Mega (E+6).
<channel>, <limit_num>, or <sense> are not specified from the given list.
A <channel> that is defined as OFF is specified.
Note that old alarm settings for a channel are lost when the function for
that channel is changed.
Alarm Limit Assignments Query
Return alarm limit data for specified channel and limit.
ALARM_LIMIT? <channel>,<limit_num>
<channel>
= (0 .. 20)
<limit_num> = 1 2
Returns HI, LO, or OFF, indicating the sense of that limit. For HI or LO
limits, the value is returned in scientific notation format with five digits of
resolution. If the channel specified is invalid, or the requested limit is not
1 or 2, an Execution Error is generated.
Remember that old alarm settings for a channel are lost when
the function for that channel is changed.
Set Date
Set instrument calendar values.
DATE <month>,<date>,<year>
<month> = (1 .. 12)
<date> = (1 .. 31)
<year> = (00 .. 99)
Invalid values generate an Execution Error.