Teledyne 235 Operating Instructions Manual page 35

235 series thermal conductivity analyzer
Table of Contents

Advertisement

Thermal Conductivity Analyzer
Refering to Figure 4-10 for positioning of the Linearizer components,
check to see that the proper breakpoint resistors are installed. Also, refer to
the Linearizer drawings that have been included in the Appendix.
The calibration curve must be known at this point and found to be non-
linear. This calibration can be done using known samples.
The curve has been studied, and breakpoint positions determined on
the most curved portions of the curve. Appropriate breakpoint resistors
have been installed.
4.4
How to Use the Linearizer
4.4.1 Nulling Amplifiers A1A
1.
Short the input (pin 3) to the common of the diff. power
supply.
2.
Connect a DVM to TP1 (testpoint 1 is located on the PC
board)
3.
Adjust trimpot P9 until the DVM reads 0 mV ±10 mV.
4.4.2 Nulling the Entire Linearizer Input to Output
1.
Maintain the shorted input of A1A. (See step 1 above.)
2.
Connect the DVM to the output of the linearizer (pin 7
of A3B).
3.
Adjust trimpot P10 until the DVM reads 0 mV ±10 mV.
4.4.3 Linearizing the Calibration Curve
1.
Remove the shorting jumper previously installed for
nulling steps.
2.
Connect a DC voltage source to the input of AlA (pin 3)
and ground. (The DVM is still connected to pin 7 of
A3B.)
3.
Apply 0 volts to the input, (V
read 0 volts as well.
4.
Make V
= V
in
test
voltage. NOTE: V
until V
= V
(1) per chart. NOTE: make sure P1
out
lin
changes V
.
out
TELEDYNE BROWN ENGINEERING
Analytical Instruments
). The output, (V
in
(1) per chart. This is the first breakpoint
must always be positive. Adjust P1
in
Linearizer 4
) must
out
4-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

238236237

Table of Contents