A-1 Specifications

Sample System: Flow indicator visible from front of unit.
Digital Interface: Full duplex RS-232 communications port.
Operating Temperature: 0-50 °C
Appendix
Packaging: General Purpose, NEMA-4, Bulkhead mount.
Sensor: Class B-1 percent analysis Micro-Fuel Cell.
Cell Block: Nylon. (Stainless steel available.)
Ranges: Three user definable ranges from 0-1 % to 0-
100 %, plus air calibration range of 0-25 %.
Autoranging with range ID output.
Positive pressure service.
Vacuum service (optional).
Auto Cal / Auto Zero. (Available with op-
tional, electrically operated valves.)
Alarms: One system-failure alarm contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Diagnostics: Start-up and on-demand, self testing function
initiated by keyboard or remote command.
Displays: 2 line by 20 alphanumeric, VFD screen, and
one 5 digit LED display.
Power: Universal power supply 110/220 VAC at 50/
60 Hz
Teledyne Analytical Instruments
Appendix
A-1