5 Maintenance
During preamp testing there is a countdown in the lower right corner
of the screen. When the testing is complete, the results are displayed.
The module is functioning properly if it is followed by OK. A number
indicates a problem in a specific area of the instrument. Refer to Table 5-1
for number-code information. The results screen alternates for a time with:
The following failure codes apply:
Table 5-1: Self Test Failure Codes
5-8
Power: OK
Analog: OK
Preamp: 2
Press Any Key
To
Continue...
Power
0
OK
1
5 V Failure
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed