It is not necessary to characterize the 5440B prior to performing the
following steps in order to meet the test uncertainty requirements for the
5790A.
3-12
the 5440B output for a null. Repeat this procedure until you obtain a null of
1 V. Set the 845A( ) OPR/ZERO to ZERO and set the 5700A to STANDBY.
Record the 5440B voltage indication in Table 3-3 after "5440B
CHARACTERIZED 6 V OUTPUT". Set the 5440B to STANDBY.
21. Set the 5700A to 2 V, OPERATE. Set the 5440B to 2 V, OPERATE. Edit the
5700A output until the Error Display matches that recorded in Table 3-2 for a
+2 V output. Set the 845A( ) range to 10 mV. Set the 845A( ) OPR/ZERO to
OPR. Reduce the 845A( ) range switch setting until the largest on-scale reading
is obtained. Adjust the 5440B for a null. Repeat this procedure until you obtain a
null of 1 V. Set the 845A( ) OPR/ZERO to ZERO and set the 5700A to
STANDBY. Record the 5440B Voltage indication in Table 3-3 after "5440B
CHARACTERIZED 2 V OUTPUT". Set the 5440B to STANDBY.
22. Set the 752A MODE switch to 1 V.
23. Set the 5440B output to the characterized 6 V output recorded in Table 3-3,
OPERATE. Set the 5700A to +0.6 V, OPERATE. Set the 845A( ) to the 10 mV
range. Reduce the 845A( ) range switch setting until the largest on-scale reading
is obtained. Adjust the 5700A for a null. Repeat this procedure until you obtain a
null of 0.1 V. Record the 5700A Error Display indication opposite +0.6 V in
Table 3-3. Set the 5700A to STANDBY.
24. Press CHNG SIGN on the 5440B. Set the 5700A to -0.6 V, OPERATE. Set the
845A( ) to 10 mV range. Repeat the nulling procedure of step 26, recording the
5700A Error Display indication opposite to -0.6 V in Table 3-3. Set the 5700A to
STANDBY.
25. Set the 5440B output to the characterized 2 V output recorded in Table 3-3,
OPERATE. Set the 5700A to +0.2 V, OPERATE. Set the 845A( ) to 10 mV
range. Repeat the nulling procedure of step 26, recording the 5700A Error
Display indication opposite to +0.2 V in Table 3-3. Set the 5700A to STANDBY.
26. Press CHNG SIGN on the 5440B. Set the 5700A to -0.2 V, OPERATE. Set the
845A( ) to the 10 mV range. Repeat the nulling procedure of step 26, recording
the 5700A Error Display indication opposite to -0.2 V in Table 3-3. Set the
5700A to STANDBY.
27. Set the 752A MODE switch to 0.1 V.
28. Repeat the procedure described in steps 26 through 29 for 5700A outputs of
+0.06, -0.06, +0.02, and -0.02 V, entering the Error Display indications in
Table 3-3.
29. Set the 5440B to output 0.6 V. Repeat the procedure described in steps 26 and 27
for 5700A outputs of +0.006 and -0.006 V entering the Error Display indications
in Table 3-3.
30. Set the 5440B to output 0.2 V. Repeat the procedure described in steps 27 and 28
for 5700A outputs of +0.002 and -0.002 V entering the Error Display indications
in Table 3-3.
31. DC source characterization is now complete. Set all outputs to standby and
remove all connections.