Fluke 5790A Service Manual page 125

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

Table 3-13. Test Record for Main Input Verification (cont.)
31
0.022
0.02
300 kHz
32
0.022
0.02
500 kHz
33
0.022
0.02
1 MHz
34
0.07
0.06
10 Hz
35
0.07
0.06
20 Hz
36
0.07
0.06
100 Hz
37
0.07
0.06
1 kHz
38
0.07
0.06
10 kHz
39
0.07
0.06
20 kHz
40
0.07
0.06
50 kHz
41
0.07
0.06
100 kHz
42
0.07
0.06
300 kHz
43
0.07
0.06
500 kHz
44
0.07
0.06
1 MHz
45
0.22
0.2
10 Hz
46
0.22
0.2
20 Hz
47
0.22
0.2
100 Hz
48
0.22
0.2
1 kHz
49
0.22
0.2
10 kHz
50
0.22
0.2
20 kHz
51
0.22
0.2
50 kHz
51A
0.22
0.2
100 kHz
52
0.22
0.2
300 kHz
53
0.22
0.2
500 kHz
54
0.22
0.2
1 MHz
55
0.7
0.6
10 Hz
56
0.7
0.6
20 Hz
57
0.7
0.6
100 Hz
58
0.7
0.6
1 kHz
59
0.7
0.6
10 kHz
[1]
720
1010
1010
[1]
720
1160
1290
[1]
720
1700
2100
230
265
265
120
145
145
60
89
90
60
89
90
60
89
90
60
89
90
120
153
163
250
302
302
500
577
577
600
760
803
600
1200
1230
200
218
218
80
92
93
30
45
46
30
45
46
30
45
46
30
45
46
65
79
79
220
260
270
250
390
420
300
970
1040
200
213
213
70
78
79
22
34
36
22
34
36
22
34
36
Calibration and Verification
Verification
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
No spec
210
82
34
34
34
34
67
No spec
No spec
No spec
210
73
27
27
27
3
3-45