Secondary Performance and Operating Characteristics (cont.)
Voltage Range
Frequency Range
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
700 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
2.2 V
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
7 V
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
22 V
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
70 V
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
Temperature Coefficient
24 Hour AC
Stability
1 C
Slow Filter
10 C to 40 C
(ppm of
Reading)
8
1.5
6
1.5
6
1.5
6
2.0
12
5.0
18
10.0
36
20.0
96
50.0
8
1.5
5
1.5
5
1.5
5
2.0
10
5.0
18
10.0
30
20.0
90
50.0
8
1.5
5
1.5
5
1.5
5
2.0
10
5.0
18
15.0
30
30.0
90
65.0
8
1.5
5
1.5
5
1.5
5
2.0
10
5.0
18
15.0
30
30.0
90
65.0
8
1.5
5
1.5
5
1.5
5
2.0
18
5.0
36
15.0
48
40.0
120
75.0
Introduction and Specifications
[1]
0 C to 10 C
Resistance
40 C to 50 C
ppm / C
3.0
3.0
3.0
3.0
8.0
10.0
20.0
50.0
3.0
3.0
3.0
3.0
8.0
10.0
20.0
50.0
3.0
3.0
3.0
3.0
8.0
15.0
30.0
65.0
3.0
3.0
3.0
3.0
8.0
15.0
30.0
65.0
3.0
3.0
3.0
3.0
8.0
15.0
40.0
75.0
1
Specifications
Input
[2]
>10 M
>10 M
50 k
50 k
50 k
1-11