Fluke 5790A Service Manual page 136

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

5790A
Service Manual
10 MHz
12 MHz
15 MHz
17 MHz
20 MHz
23 MHz
26 MHz
28 MHz
30 MHz
1
7 V
2
7 V
3
7 V
4
7 V
5
7 V
6
7 V
7
7 V
8
7 V
9
7 V
10
7 V
11
7 V
12
7 V
13
7 V
14
7 V
3-56
Table 3-15. Wideband Flatness Verification Worksheet (cont.)
Table 3-16. Wideband Verification Test Record
3.2 V
10 Hz
0.012
3.2 V
20 Hz
0.007
3.2 V
50 Hz
0.007
3.2 V
100 Hz
0.007
3.2 V
200 Hz
0.007
3.2 V
1 kHz
0.08
3.2 V
2 kHz
0.007
3.2 V
10 kHz
0.007
3.2 V
20 kHz
0.007
3.2 V
50 kHz
0.007
3.2 V
100 kHz
0.007
3.2 V
200 kHz
0.020
3.2 V
500 kHz
0.020
3.2 V
700 kHz
0.025
___________Range
[1]
0.10
[1]
0.10
[1]
0.03
[1]
0.03
[1]
0.03
[2]
N/A
[1]
0.03
[1]
0.03
[1]
0.03
[1]
0.03
[1]
0.03
[1]
0.03
[1]
0.03
[1]
0.05
X
X
X
X
X
X
X
X
X
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
0.31
No Spec