Calibration Path; Exploded View Of Scale - Health O Meter PRO PLUS 2450KL Operation Manual

Portable wheelchair scale
Table of Contents

Advertisement

CALIBRATION PATH

The calibration is performed in Kg or Lb, according to the units of measure used upon entering
into the calibration path.
1.
Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
2.
Press ENTER
3.
Using
keys and/or the keypad, set the
calibration load to at least 200 lbs (for best results
and greater accuracy, use a 600 lb weight). Press
ENTER
4.
Please clear the scale platform and press ENTER
5.
Please wait without touching or shaking the scale
until the zero calibration process is finished
6.
Load the scale platform with the required weight
and press ENTER
7.
Please wait without touching or shaking the scale
until the calibration process is finished
8.
Press ENTER
9.
Please remove weight from the scale platform
10. Please wait until the scale resumes normal
operation
Operator Action

EXPLODED VIEW OF SCALE

Calibration
Process
Enter load weight
600.0
Zero calibration
Please clear the scale
Zero calibration
Please wait
Weight calibration
Put: XXX.X
Weight calibration
Please wait
Calibration
factor: X.XXXXX
Calibration
Please clear the scale
Rebooting
Please wait
Health O Meter
Pro Plus
19