Model 2101KL
Digital Platform Scale

User Instructions

P/N UM2101KLE_ Rev20170710
1

Advertisement

Table of Contents
loading

Summary of Contents for Health O Meter 2101KL

  • Page 1: User Instructions

    Model 2101KL Digital Platform Scale User Instructions P/N UM2101KLE_ Rev20170710...
  • Page 2: Table Of Contents

    Model 2101KL ® Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Cautions and Warnings ..................... 3 Specifications ......................4 Certifications / Connectivity / Disposal ..............5 Assembly Instructions ....................
  • Page 3: Cautions And Warnings

    Using the Scale with the Optional AC Adapter  ® Operate this scale exclusively with the AC adapter provided by Health o meter Professional. Use of an unspecified adapter will void the warranty and can pose a serious safety hazard.
  • Page 4: Specifications

    SPECIFICATIONS General Model 2101KL digital scales use highly sophisticated microprocessor technology. Each precision scale is designed to provide accurate, reliable and repeatable weight measurements. This scale uses motion-sensing weighing technology in order to determine the actual weight of a moving patient.
  • Page 5: Certifications / Connectivity / Disposal

    To enable reliable transmission of weight, height, and Body Mass Index (BMI) data, this scale is designed to interface with a computer, EMR software or other electronic device ® via various Health o meter Professional Connectivity Solutions. This connectivity helps minimize transcription errors when recording patient data.
  • Page 6: Assembly Instructions

    Carefully inspect the carton for shipping damage before unpacking. If damage is found, contact ® your shipper or a Health o meter Professional representative immediately at 1-800-815-6615. Claims must be filed with the shipper as soon as possible after receipt of the package. The following information details what you will find inside the main carton as you unpack the parts for assembly.
  • Page 7 ASSEMBLY INSTRUCTIONS (CONT.) Health o meter® Professional Scales recommends that assembly is performed by two people. Warning: Do not drop scale, as this can permanently damage the load cells. Caution: Do not insert the batteries into the display head until it is mounted onto the scale.
  • Page 8 ASSEMBLY INSTRUCTIONS (CONT.) 5. Pull forward on the side handrails to tip the scale and gently place it face down on the floor. Grasp the load cell cable protruding from the bottom of the platform and feed it through the protective metal shield.
  • Page 9 ASSEMBLY INSTRUCTIONS (CONT.) Route the load cell cable to the side and then between the washer and the silver lip in the bracket. Clip the load cell cable into the divot in the middle of the cover. Obtain the plastic bracket cover and the two remaining M4 x 0.7 x 10mm machine screws.
  • Page 10: Set-Up Instructions

    AC adapter into the power source. Warning: To avoid a safety hazard, do not use an unspecified AC adapter. Only use the Health o meter® Professional AC adapter model ADPT30 designed for this scale. Power Adapter Jack For battery operation, see Battery Installation below.
  • Page 11 SET-UP INSTRUCTIONS (CONT) The scale must be powered on before performing any of the following operations. Press to turn on the scale. Unit of Measure Configuration Unit The weight measuring unit (pounds/LB or kilograms/KG) can be changed by pressing To lock the weight measuring unit, follow this procedure. Help 1.
  • Page 12 After each use, the scale will clear the weight of the patient and set itself back to the initial ZERO ® level it found on start-up. Health o meter Professional recommends keeping this setting as Enable. Note: Hitting a key to wake the scale from sleep mode will not prompt the scale to zero.
  • Page 13: Operating Instructions

    OPERATING INSTRUCTIONS and selecting “Key Definitions” Help Key definitions can also be found by pressing in the submenu. Keypad Function Description Powers scale on and off. Zeroes the scale prior to a weighing. Also used as a screen Zero/ navigation key to exit a display menu. (Clear) Holds the weight value on the display until the button is pressed Hold/...
  • Page 14 OPERATING INSTRUCTIONS Weighing a Patient 1. Make sure there is nothing on the weighing platform and the display shows “0.0”. 2. Ask the patient to step on the scale. The scale will begin to display the patient’s weight and the word “PROCESSING”. Depending on the movement of the patient on the scale, it may take several seconds for the scale to lock onto the weight.
  • Page 15 OPERATING INSTRUCTIONS (CONT) BMI Function 1. Make sure there is nothing on the weighing platform and the display shows “0.0”. 2. Ask the patient to step on the scale. The scale will begin to display the patient’s weight and will lock on the patient’s weight when completed.
  • Page 16 OPERATING INSTRUCTIONS (CONT) Tare Function When using this scale, the weight of an object, such as a portable oxygen tank, can be subtracted from the total weight in order to determine the weight of the patient alone. The Tare Function automatically performs this subtraction. The scale allows a tared weight to be entered manually by using the keypad, or automatically, as outlined in this section.
  • Page 17 OPERATING INSTRUCTIONS (CONT) To Tare with a Known Weight / Pre-Tare (Limit 200 lb / 91kg) 1. Wait until “0.0” appears on the display. Tare 2. Press and hold for approximately 3 seconds. The display will change and prompt entry of the weight value to be tared.
  • Page 18 OPERATING INSTRUCTIONS (CONT) Parent/Child Function ® The Parent/Child function on Health o meter Professional scales provides the convenience of weighing a small child or infant while being held by a parent. Follow these simple steps to subtract the parent’s weight from the total so only the child’s weight is displayed.
  • Page 19: Maintenance

    Maintenance The following pages provide instructions for maintenance, cleaning, calibration, and troubleshooting for the 2101KL scales. Maintenance operations other than those described in this manual should be performed by qualified service personnel. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function.
  • Page 20: Calibration

    This scale has been factory calibrated, and does not require calibration prior to use. Calibration of your 2101KL scale can be performed using calibrated kilogram (kg) or pound (lb) weight standards. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is 1000 lb / 454 kg.
  • Page 21 CALIBRATION (CONT) 7. The display will prompt for the calibration weights to be placed on the platform. Press after calibration Enter weights are placed on the platform. 8. The display will show “Calibrating” during the calibration process. 9. If calibration is successful, the display will show “Calibration Complete”...
  • Page 22: Troubleshooting

    TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to ® correct any failures. For further assistance, contact Health o meter Professional Scales Technical Support at 1-800-638-3722. Symptom Possible Cause Corrective Action 1. Dead Battery 1.
  • Page 23: Warranty

    ® Health o meter Professional products are manufactured, designed, and owned by Pelstar, LLC. ® We reserve the right to improve, enhance, or modify Health o meter Professional product features or specifications without notice. © Pelstar, LLC 2017 P/N UM2101KLE_ Rev20170710...