Troubleshooting

11.10 Status alarms

Status alarm can be viewed with ProLink II. A list of status alarms and possible remedies is provided
in Table 11-4.
Table 11-4
Status alarms and remedies
Alarm
code
ProLink II label
A001
CP EEPROM Failure
A002
CP RAM Failure
A003
Sensor Failure
A004
Temp Out of Range
A005
Mass Flow Overrange
A006
Characterize Meter
A008
Density Out of Range
A009
Xmtr Initializing
A010
Calibration Failure
Configuration and Use Manual
Possible remedy
Cycle power to the flowmeter.
The flowmeter might need service. Contact Micro Motion. See Section 1.8.
Cycle power to the flowmeter.
The flowmeter might need service. Contact Micro Motion. See Section 1.8.
Check the test points. See Section 11.23.
Check the sensor coils. See Section 11.25.
Check wiring to sensor. See Section 11.14.2.
Check for slug flow. See Section 11.17.
Check sensor tubes.
Check the test points. See Section 11.23.
Check the sensor RTD reading(s). See Section 11.25.
Check wiring to sensor. See Section 11.14.2.
Verify flowmeter characterization. See Section 4.2.
Verify that process temperature is within range of sensor and transmitter.
Contact Micro Motion. See Section 1.8.
Check the test points. See Section 11.23.
Check the sensor coils. See Section 11.25.
Verify process.
Make sure that the appropriate measurement unit is configured. See
Section 11.19.
Verify 4 mA and 20 mA values. See Section 11.20.
Verify calibration factors in transmitter configuration. See Section 4.2.
Re-zero the transmitter.
Check the characterization. Specifically, verify the FCF and K1 values. See
Section 4.2.
If the problem persists, contact Micro Motion. See Section 1.8.
Check the test points. See Section 11.23.
Check the sensor coils. See Section 11.25.
Verify process. Check for air in the flow tubes, tubes not filled, foreign material
in tubes, or coating in tubes.
Verify calibration factors in transmitter configuration. See Section 4.2.
Perform density calibration. See Section 10.5.
Allow the flowmeter to warm up. The error should disappear once the flowmeter
is ready for normal operation.
If alarm does not clear, make sure that the sensor is completely full or
completely empty. Verify sensor configuration and wiring to sensor.
If alarm appears during a transmitter zero, ensure that there is no flow through
the sensor, then retry.
Cycle power to the flowmeter, then retry.