Rosemount ™ Cx1100 In-Situ Oxygen Transmitter System Alarms - Emerson Rosemount CX1100 Reference Manual

In-situ oxygen transmitter
Hide thumbs Also See for Rosemount CX1100:
Table of Contents

Advertisement

Parameters
May 2023
Table 7-2: Probe Parameters
Parameter label
Calibration Log | Constant
Calibration Log | Impedance
Calibration Log | Log Time
Cal Result
7.2 
Rosemount
system alarms
Table 7-3: System Alarms
Alarm label
NV Memory Fail
SB Disconnect
SB Mismatch
SB Version
SB Bootload Mode
Table 7-4: Probe Alarms
Alarm label
O
T/C Reversed
2
42
(continued)
CX1100 In-Situ Oxygen Transmitter
Help messages and recommended actions
Checksum error was detected in the nonvolatile memory
configuration data when the unit was turned on. Default
values have been loaded. Check to see that your
configurations have not been changed. Cycle the power to
clear alarm.
Communication failure detected between the sensor board
and the main board, indicating the sensor board has been
disconnected from the main board.Check the connecting
cable pins.
The sensor type does not match, indicating wrong sensor
board is connected.
The sensor board software version does not match the
expected version, indicating the sensor board software
needs to be upgraded. Place the jumper on J1 of the
sensor board and cycle the power to start the sensor board
software upgrade.
The Bootload Mode jumper at J1 of the sensor board is on
at power up. This mode should only be used to upgrade the
sensor board software. Remove the jumper from J1 of the
sensor board and cycle the power to clear the alarm.
Help messages and recommended actions
The sensor heater thermocouple voltage is reading
a negative voltage, indicating the thermocouple wire
connections may be reversed. Check wiring.
Description
Previous calibration constant. there are ten
calibration results.
is the most recent, and
1
is the least recent calibration constant.
Previous cell impedance. This is the sensor
resistance that was calculated as a result of
previous successful calibration. There are ten
calibration results.
Index 1
and
is the least recent sensor
Index 10
resistance measured.
Time stamp of the previous successful
calibration. There are ten calibration time
stamps.
is the most recent, and
Index 1
is the least recent time stamp.
Index 10
Calibration result
Reference Manual
00809-0100-4110
10
is the most recent,
Rosemount CX1100