Hanna Instruments HI 720 Instruction Manual page 12

Conductivity process controller with inductive probe
Hide thumbs Also See for HI 720:
Table of Contents

Advertisement

Setup Item
Description
INPUT
I.00
Input selection
(note 10)
I.03
Digital transmitter
address (active if I.00=Digital transmitter)
I.04
Digital transmitter
baud rate
I.10
Calibration timeout
I.11
Life check
(see notes 9, 20)
REAL TIME CLOCK (see Note 4)
r.00
Current day
r.01
Current month
r.02
Current year
r.03
Current time
COMMUNICATION
P .00
Connection type
P .01
Baud rate
P .02
Call answer enable
P .03
Modem country
code
P .11
PIN code
P .12
Phone number #1
(see note 21)
P .13
Phone number #2
(see note 21)
22
Valid Values
Default
Inductive probe or
Inductive
Digital transmitter
probe
00 to 99
00
1200, 2400, 4800,
19200 bps
9600 or 19200 bps
0 to 12 months
12 months
Disabled,
Disabled
1 hour, 2 hours,
or 4 hours
01 to 31
from RTC
01 to 12
from RTC
2000 to 2099
from RTC
00:00 to 23:59
from RTC
PC,
PC
HI 504900 GSM module,
HI 504901 GSM supervisor,
HI 504902 Modem
1200, 2400, 4800,
19200 bps
9600 or 19200 bps
On, Off
Off
Insert the phone
000
country code for the
HI 504902 modem
(see note 35)
0000 to 9999
0000
+ --------------------
("Not set" means no number entered)
+ --------------------
("Not set" means no number entered)
Setup Item
Description
COMMUNICATION
P .14
Remaining SMS
(see note 22)
P .15
Repeated SMS
P .16
SMS delay between
messages (see note 24)
P .17
SIM card expiration
day (see note 25)
P .18
SIM card expiration
month (see note 25)
P .19
SIM card expiration
year (see note 25)
CLEANING
Simple Cleaning
L.00
Rinsing time
L.01
Pause time
Advanced Cleaning (see Note 19)
L.10
Pre-rinsing time
L.11
Cleaning time
L.12
Rinsing time
L.13
Pause time
L.14
Minimum pause time 10 to L.13 minutes
L.15
Cleaning trigger
L.16
Repeated cycles no.
L.17
Economy cycle no.
ERROR CONFIGURATION (see Note 5)
E.00
High alarm (00)
Valid Values
Default
000 to 200, or 222
100
0 to 5 (see note 23)
2
05 to 60 minutes
10 minutes
01 to 31
01
01 to 31
01
2000 to 2099
2010
5 to 99 seconds
20 seconds
10 to 9999 minutes
1440 min.
0 to 99 seconds
20 seconds
0 to 99 seconds
10 seconds
5 to 99 seconds
20 seconds
10 to 9999 minutes
1440 min.
10 minutes
Timer only,
Timer only
External only,
Timer and external,
Timer masked by external
0 to 10
0
0 to 10
0
Alarm relay -------------> On
22 mA fault current----> On
3.6 mA fault current---> Off
Auto cleaning ----------> Off
SMS sending -----------> Off
23

Advertisement

Table of Contents
loading

Table of Contents