C. Replacing the Fuses
Warning
To avoid electrical shock or arc blast, or personal injury or damage to the
Meter, use specified fuses ONLY in accordance with the following procedure.
To replace the Meter's fuse:
1.
Turn the Meter to OFF position.
2.
Remove the screw from case bottom, and separate the case bottom from the
case top.
3.
Remove the fuse by gently prying one end loose, and then take out the fuse
from its bracket.
4.
Install ONLY replacement fuses with the identical type and specification as
follows and make sure the fuse is fixed firmly in the bracket.
Fuse: 0.2A, 250V, fast type, 5x20mm.