The pressure is automatically heightened to the level for
measurement of the blood pressure based on the last-mea-
sured systolic blood pressure.
At the first measurement after battery replacement the pres-
sure is heightened to 160mmHg.
If the level of the pressure is not enough, the pressure is
automatically heightened to the level for measurement of
the blood pressure again.
When the "POWER / START" switch pressing, the pressure
level for measurement will change at optional level over
160 mmHg.
But if the "POWER / START" switch is released before 160
mmHg, the pressure is heightened to 160mmHg.
SYSTOLIC mmHg
When the pressure lowers by 15mmHg after pressuriza-
tion stops, the mark of "Q" is displayed.
SYSTOLIC mmHg
If pulse is recognized, the pressure is shifted to the dias-
tolic blood pressure indicating section and change over,
synchronized with the pulse.
- Eng 8 -
DIASTOLIC mmHg
PULSE /min
DIASTOLIC mmHg
PULSE /min