Measuring method: | Oscillometric system |
Indication: | Digital display, Height: 19mm |
Measuring range: | Pressure; 0 ~ 280mmHg, Pulse; 40 ~ 200 beat/min. |
Accuracy: | Pressure; ±3mmHg, Pulse; ±5% |
Inflation: | Automatic inflation by internal pump |
Pressurization setting: | 4 stage: 130, 160, 200, 250mmHg |
Deflation: | Automatic constant speed deflation system |
Exhaust: | Quick exhaust by electromagnetic valve |
Memory function: | 7 data |
Power source: | 4 "AA" (R6P, LR6) batteries |
Battery life: | Approx. 400 times/LR6 Approx. 200 times/R6P (once a day, 160mmHg, room temp. 22°C) |
Rated voltage: | 6V DC |
Working temp. & humidity: | +10°C +40°C, 30% 85%RH |
Storage temp. & humidity: | –20°C +60°C, 10% 95%RH |
Outside dimensions: | 135mm(W) × 130mm(D) × 50.5mm(H) |
Weight: | Approx. 360g (Apparatus and batteries) |
Electric shock protection: | Type B equipment |
Accessories: | Cuff, 4 "AA" (R6P) batteries, Instruction manual |
*Specifications may be changed without notice in the event of improvements being made.
After cleaning the cuff with a neutral detergent, lightly rinse it with water and let it dry in the air. Hard rubbing or washing in a washing machine will cause air leakage. Also, be careful not to get water into the air hose.
General Remarks on blood pressure measurement
For proper measurement of blood pressure
Precautions for use
Precautions for maintenance
Nature and frequency of maintenance:
This products is designed for use over an extended period of time; however, it is generally recommended that it be inspected every two years to ensure proper function and performance.
Protect the nature environment:
Please help protect natural environment by respecting national and/or local recycling regulations when disposing of the batteries and the products at the end of their useful lives.
Reference European standard:
This products complies with EC instructions 93/42/EEC (Medical Device Directive) and with the following standards: EN 1060-1: 1995 "Noninvasive Sphygmomanometers, Part 1 General Requirements" and EN 1060-3: 1997 "Noninvasive Sphygmo-manometers, Part 3 Supplementary Requirements for Electromechanical Blood-Pressure Measuring Systems".
Manufacturer:
Name: Premier Precision Ltd.
Address: Block E, G-H, 6 Phase 2, Yip Fat Factory Building, 73-75 Hoi Yuen Road, Kwun Tong, Kowloon, Hong Kong
European Representative:
Name: WvW elektronische Geräte Vertrieb GmbH.
Address: Robert-Koch-Straße 20, D-80538 München, GERMANY
Insert the Hose plug from the cuff into the hose socket at the mainframe until it is tightly connected.
Press the "Power switch" for power ON.
All display marks appear for approx. 1 second and buzzer beeps.
After all display marks are disappeared, deflation mark " " appears for approx. 1 second (three short beeps notify you of ready for measurement) then the last pressure setting value appears for approx. 3 seconds. Then the " 0 " appears at the display.
Ready for measurement
Unfasten and open the cuff as figure.
Wind the arm cuff around the left upper arm tightly so that the bottom edge of the cuff is positioned about 1-2 cm (about 1 inch) above the elbow joint.
The cuff should be wound so that one finger can be inserted under the cuff. Note that uneven cuff winding will prevent proper measurements.
Rest the elbow on a table so that the section of the upper arm is at the height of the heart. Relax the lower arm across the table and remain still. Make measurements in either a seated or reclining posture.
Press the "Pressure Setting Switch" and set the value to a value approximately 20~30mmHg above your normal systolic blood pressure.
There are 4 settings; 130, 160, 200 and 250mmHg.
After press on the "Pressure Setting Switch", the set pressure value will be displayed at the left and the right side of the display.
Press the "Start Switch".
The arm cuff pressurizes when the "Start Switch" is pressed.
When the pressurization is reached to the set value, air inflation is stopped automatically.
When pressurization is completed, the automatic exhaust mechanism will gradually reduce arm cuff pressure, and the mark " " appears indicating that measurement is in progress.
When a pulse is detected, the pressure value moves to the right side of the display. At this time the mark " " flashes and buzzer sounds.
When the diastolic pressure value is detected, the measurement is completed and buzzer sounds.
Systolic pressure is displayed on the left and diastolic pressure on the right side of the display.
2 to 3 seconds after completion of measurements, the mark " " appears and air will deflate immediately.
2 or 3 seconds later, pressure value and pulse appear alternately on the display.
Press the "Power Switch" to turn OFF.
When a measurement need to be made again, just press the "Start Switch" again.
When the pressure setting is needed to be changed, press the "Pressure Setting Switch" and select optimum value.
This blood pressure monitor has a function which is able to display last measured value.
If this unit is left on after measurement, an Automatic Power- OFF Function turns the power OFF after approximately 3 minutes. To continue measurement press the power switch to turn the unit ON.
If the batteries become weak, the " " mark will appear in the display, indicating that batteries should be replaced with new ones.
Remove the battery cover at mainframe and replace the new four pieces of batteries.
Type "AA" (R6P, LR6) battery or equivalent is suitable for this blood pressure monitor.
DISPLAY MARK | CONDITION/CAUSE | CORRECTIVE ACTION |
Measurement is begun, but the pressure was insufficient. | Change the pressurization setting one level by pressing "Pressure Setting Switch" and measure once again. (Automatic re-measuring will be done only 1 time.) | |
Appears when the blood pressure could not be measured accurately. (Some noise was detected or moved arm during measuring.) | Repeat measurement again. First exhaust air from the cuff, reapply the cuff properly and remeasure. | |
Appears when the pulse could not accurately be measured. | ||
Appears when the batteries become weak. | Replace the batteries. | |
Appears when no memory is stored at pressing of "Memory Recall Switch" | Take a measurement. Blood pressure will store automatically. | |
Units has a trouble. | Ask purchased shop. |
Consumer Electronics Division
CBM Bldg. 5F, 5-68-10, Nakano, Nakano-ku, Tokyo 164-0001, Japan
Phone: (03)5345-7430 Fax: (03)5345-7431 E-mail: info-con@jcbm.co.jp
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the CH-403C that isn't answered in the manual? Leave your question here.