FUNC

Syntax
function
max._input
Function. Selects the type of measurement (AC voltage, DC current. etc.). lt also
allows you to specify the measurement range and resolution. (The FUNC header
is optional and may be omitted.)
FUNC [function][,max._input][,%_resolution]
or
[FUNC] function[,max._input][,%_resolution]
The function parameter designates the type of measurement. The parameter
choices are:
Numeric
function
Query
Parameter
Equiv.
DCV
1
ACV
2
ACDCV
3
OHM
4
OHMF
5
DCI
6
ACI
7
ACDCI
8
FREQ *
9
PER *
10
DSAC *
11
DSDC *
12
SSAC *
13
SSDC *
14
*These functions require additional explanation and are documented individually in
this chapter. Refer to the corresponding DSAC, DSDC, FREQ, PER, SSAC, or
SSDC command for details.
Power-on function = DCV.
Default function = DCV.
Selects a fixed range or the autorange mode. To select a fixed range, you specify
max._input as the absolute value (no negative numbers) of the input signal's
maximum expected amplitude (or the maximum resistance for ohms
measurements). The multimeter then selects the correct range.
Description
Selects DC voltage measurements
Selects AC voltage measurements (the mode is set by the
SETACV command)
Selects AC+DC voltage measurements (the mode is set
by the SETACV command)
Selects 2-wire ohms measurements
Selects 4-wire ohms measurements
Selects DC current measurements
Selects AC current measurements
Selects AC+DC current measurements
Selects frequency measurements
Selects period measurements
Direct sampling, AC coupled
Direct sampling, DC coupled
Sub-sampling, AC coupled
Sub-sampling, DC coupled
Chapter 6 Command Reference

FUNC

183