Omron ZG2 Series User Manual page 104

Smart sensor 2d profile measuring sensors
Hide thumbs Also See for ZG2 Series:
Table of Contents

Advertisement

MEAS/AVERAGE
[1]
OK
 FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[AVERAGE]
Setting value
1, 2, 4, 8, 16, 32, 64
Important
When measuring edge position and edge width, set the average number of times to "1". Setting
to a value other than "1" will result in a drop in the measurement accuracy.
Note
Averaging is not performed for the data on the right edge because the required number of
samples cannot be obtained.
MEAS/AVERAGE
[ 1]
OK
Setting the smoothing function
Changes in data are smoothed out using the intermediate values of adjacent data.
Smoothing is performed in the "width direction (X-axis direction)." This setting is effective
in filtering noise such as spikes.
MEAS/SMOOTH
[LOW]
OK
 FUN mode-[MEAS]-[IMAGE]-[PROFILE]-[SMOOTH]
Setting value
OFF, LOW, MID, HIGH
ZG2 User's Manual
MEAS/AVERAGE
[64]
OK
Description
Sets the number of data to average.
(default value: 1)
[64]
OK
MEAS/SMOOTH
[HIGH]
OK
Description
Sets the smoothing strength. (default value: LOW)
Image Adjustment
101

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zg2-wds70Zg2-wds22Zg2-wds8tZg2-wds3vt

Table of Contents