To set the sensitivity for the external sensor:
1. Turn on the Product while pushing  at the same
time.
The screen shows the model number and software
version number.
2. Push  again to enter the sensitivity setting.
The screen below shows that the initial value is 10.00,
which means 10.00 mV/m/s².
You must set the sensitivity in mV/m/s² units.
3. Push  () or  () to enter setting mode. 10.00
flashes.
4. In setting mode (when the reading flashes), push  to
increase 0.01, or push  to decrease 0.01. The unit is
mV/m/s².
5. When you get the required sensitivity, push  to
save the setting and exit setting mode.
The external sensor can be purchased separately.
Set the sensitivity according to the reference data
that is shipped with the external sensor.