Suregri tapered temperature probe

Advertisement

Quick Links

To avoid electrical shock, do not use this
probe when voltages exceeding 24 V ac rms
or 60 V dc are present. The probe tip is
electronically connected to the output
terminals.
Introduction
The 80PK-26 SureGrip
essentially a bead probe with a handle that can be used
as a surface probe. The stainless steel sheath allows the
bead to measure temperatures higher than bead probes
with flexible leads. The handle also allows you to apply
more focused surface pressure without having your hand
in close proximity to the source. The 40-inch (1-meter)
cable is terminated with a Type K miniature thermocouple
connector with 0.792-mm (.312-in) pin spacing. The
80PK-26 can be used with any temperature-measuring
instrument that is designed to accept Type K
thermocouples and has miniature connector input.

Specifications

Type
K Standard Grade Ni-Cr vs Ni-Al (Chromel vs Alumel)

Measurement Range

-40 °C to 816 °C (-40 °F to 1500 °F)
Accuracy
(With respect to ANSI MC96.1-1982-Standard Limits of
Error):
All error calculations should be done in
scaled to
Range
-40 °C to 0 °C
(-40 °F to 32 °F)
0 °C to 275 °C
(32 °F to 527 °F)
275 °C to 816 °C
(527 °F to 1500 °F)
November 2004
© 2004 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies
SureGrip

Temperature Probe

XWWarning
TM
Tapered Temperature Probe is
Note
°
F.
80PK-26
Tapered
TM
Instruction Sheet
°
Accuracy
(% of reading)
±2.2 °C
±1.1 °C
±0.4 %
®
C, then

Advertisement

loading

Summary of Contents for Fluke 80PK-26

  • Page 1: Temperature Probe

    The 40-inch (1-meter) cable is terminated with a Type K miniature thermocouple connector with 0.792-mm (.312-in) pin spacing. The 80PK-26 can be used with any temperature-measuring instrument that is designed to accept Type K thermocouples and has miniature connector input.
  • Page 2 Output @ 25 °C (77 °F) = 1.00 mV (reference junction @ 0 °C) Seebeck Coefficient 25 °C (77 °F) = 40.50 µV / °C Measurement Time (Time Constant): 3.0 seconds for air moving at 3.33 meters / second (10.9 feet / second) at sea level pressure (5 time constants = 1 complete step change, i.e., 15 seconds) Maximum Voltage...
  • Page 3: Measurement Considerations

    Even though the sheath to junction transition is sealed, it is not recommended to immerse the probe in liquid or saturated vapor. The 80PK-26 should not be used in reducing atmospheres or in a vacuum. Excess mounting force can deform or crush the bead. With the probe perpendicular to the measurement surface, do not apply more than 12.25 new tons (2.75 pounds-force).
  • Page 4: Troubleshooting

    Check for an open circuit indicator on the measuring instrument. Some temperature measuring instruments have a built-in circuit to indicate if the connected probe is open. (All Fluke Temperature-measuring instruments have this feature.) Refer to the owner’s manual accompanying the measuring instrument to see if this feature is available.
  • Page 5: Contacting Fluke

    For application or operation assistance, or information on Fluke products, call: USA: 1-888-44-FLUKE (1-888-443-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 For USA Service: 1-888-99-FLUKE (1-888-993-5853) Or, visit Fluke's Web site at www.fluke.com. To register your product, visit register.fluke.com...
  • Page 6 ONE YEAR LIMITED WARRANTY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.